Intended use :
This equipment provides a safety heating and stabilized temperature within the range from slightly above to
49.9ºC. Ideal for chemical, biological, clinical and laboratory applications. It is also applied to desfrost the frozen
blood products including plasma and other medical, biological and scientific products. Clinical purpose: Thawing
liquefaction of the frozen plasma at controlled temperature of +37ºC±0.5ºC and controlled time duration thus
maintaining the properties of clotting factors in the plasma.
Specifications : | |
Model | YCW-06S |
Overall dimension (±10mm) | 690 x 450 x 460mm approx. |
Basket dimension (±5mm) | 140 x 90 x 250mm |
Maximum capacity | 6 or 8 bags, 250ml ~ 500ml each |
Gross weight | 31 kg |
Shaking stroke | 20mm reciprocal per cycle |
Shaking speed | Fixed at 80 ~ 100 rpm |
Defrost temperature | 34ºC ~ 37ºC |
Defrost time | 15 ~ 20 minutes from the 36.5ºC |
Ambient temperature range | 15ºC ~ 32ºC (Do not exceed 37ºC) |
Working range | Ambient ~ 49ºC |
Temperature accuracy | ±0.1ºC at 36.5ºC |
Temperature stability | ±0.1ºC at 36.5ºC |
Temperature resolution | 0.1ºC |
Timer | Digital 99 minutes adjustable |
Temperature control | PID |
Temperature sensor | PT-100 |
Water circulation | Yes, water stirrer |
Water drain | Yes, |
Overheat alarm | Yes, alarm + cutout |
Sensor error alarm | Yes alarm (Er 2) |
Double safety device | Relay to shut out when overheating |
Low water level alarm | Yes alarm (Er 1) |
Heating element | 1000W. Stainless steel sheathed, |
Power source | 115V/60 Hz, or 230V/50Hz; |
Standard accessory | Gable Lid, Basket x6, Safety Bag x100 |