Horizontal Laminar Air Flow, Model: CLB-202D 04

It is a horizontal airflow type that the clean air filtered through the HEPA filter flows from the inside the chamber to the direction of the front entrance, and the inflow of the outside air which can contaminate the specimen is blocked as the inside the chamber is maintained as positive pressure (+) without the front door.

Clean work space at any points inside the chamber is available, and if any contaminated particles are generated, they are quickly eliminated. It is mainly used for the work of blood agar plate, experiment or work under the aseptic, non-vibration environment.

General specification

Dimension
standard Exterior W / mm D / mm H / mm
1340 745 1300
Interior W / mm D / mm H / mm
1219 560 584
Color Gray
Door Acrylic panel door
Control panel LCD digital display panel
Material
Exterior Steel plate cold commerica / Aluminium
Interior Stainless steel 304
Support stand Carbon steel pipes for ordinary piping
Performance
Certification CE
Avg. downflow velocity (m/s) 0.3~0.4 (variable set speed)
Filter efficeincy HEPA filter (>99.99% at 0.3㎛) / Class 100
Pre-Filter Nylon filter
Noise level (db) ≤65db(A)
Safety alarm system X
Electrical
Power supply 220V, 50/60Hz (110V is available)
FL lamp 36W x 1EA
UV lamp 30W x 1EA
Power Receptacle 2EA
Valve
Utility Valve Option 1EA (vacuum)
Option
Support stand (height) / Frame – 633mm
Leveling caster (height) – 71mm
Etc
Differential pressure gauge 1EA